SR-50L Constant temperature and humidity test chamber.

Talk to an Expert
Category: Tag:

SR-50L Constant temperature and humidity test chamber main parameter

1.Internal box dimensions: W350H400D350mm
2.External box dimensions: W600H1450D1000mm
2.Temperature range: -60°C to 150°C
3.Temperature fluctuation: ±0.5°C
4.Temperature deviation: ≤±2°C
5.Heating-up time: It takes about 40 minutes to rise from room temperature to 150°C (non-linear under no-load condition)
6.Cooling-down time: It takes about 60 minutes to drop from room temperature to -60°C (non-linear under no-load condition)
7 .Humidity range: 20% to 98%R.H.
8.Humidity fluctuation: ±3%R.H.
9.Humidity deviation: ≤3%

SINRAD tech Temperature and humidity test chamber with a 7-inch display LCD touch panel, with a dialogue-style input of data on the screen. Temperature and humidity can be programmed simultaneously, and curves are displayed, including the set value/display value curves. It can display various alarms separately. When a fault occurs, the fault can be displayed on the screen, and the fault can be eliminated, as well as the misoperation. It has multiple groups of PID control functions and precise monitoring functions, and the data is displayed on the screen.
  1. Controller specifications:
    a. Precision: Temperature ±0.1℃, Humidity ±1%RH
    b. Resolution: Temperature ±0.1, Humidity 1%R.H.
    c. Temperature slope: Can be set from 0.1 to 9.9
    d. Has upper and lower limit standby and alarm functions
    e. Temperature and humidity input signal selection: PT 100Ω×2 (dry bulb and wet bulb)
    f. Temperature and humidity conversion output: 4~20MA
    g. 6 groups of P.I.D control parameter settings, P.I.D automatic calculation, automatic calibration of dry and wet bulbs
  2. Screen display functions:
    a. Touch input, with options directly displayed on the screen
    b. The temperature and humidity set value (SV) and the actual value (PV) are directly displayed (in English)
    c. Can display the current running program number, segment number, remaining time and number of cycles
    d. Function of cumulative running time
    e. The temperature and humidity program set values are displayed as graphic curves, with the function of real-time display of the program curve execution
    f. Has a separate program editing screen, allowing direct input of temperature, humidity and time
  3. Program capacity and control functions:
    a. Available program groups: Up to 100 groups at most
    b. Available memory capacity: A total of 5000 SEGMENTS
    c. Repeatable execution commands: Each command can be executed up to 9999 times
    d. The program is made in a dialogue style, with functions such as editing and clearing
    e. SEGMENTS time setting: 0~530 hours and 59 minutes
    f. Has the function of program memory during power failure, and will automatically start and continue to execute the program after power restoration
    g. The graphic curve can be displayed in real time during program execution
    h. Has the function of date and time adjustment
    i. Button and screen lock (LOCK) function

VIdeo Tutorial for SR-50L Constant temperature and humidity test chamber.

 